Theory and Design for Mechanical Measurements
6th Edition
ISBN: 9781118881279
Author: Richard S. Figliola, Donald E. Beasley
Publisher: WILEY
expand_more
expand_more
format_list_bulleted
Textbook Question
Chapter 5, Problem 5.36P
A pressure measuring system outputs a voltage that is proportional to pressure. It is calibrated against a transducer standard (certified error: within ±0.5 psi) over its 0-100-psi range with the results given below. The voltage is measured with a voltmeter (instrument error: within ±10 pV; resolution: 1 pV). The engineer intending to use this system estimates that installation effects can cause the indicated pressure to be off by another ±0.5 psi. Estimate the expanded uncertainty at 95% confidence in using the system based on the known information.
£lmVJ: 0.(X)4 0.399 0.771 1.624 2.147 4.121
p[psi]: 0.1 10.2 19.5 40.5 51.2 99.6
Expert Solution & Answer
Want to see the full answer?
Check out a sample textbook solutionStudents have asked these similar questions
The transducer specified in Table 1.1 is chosen to measurea nominal pressure of 500 cm H2O. The ambient temperature is expected to vary between 18 ∘C and 25 ∘C duringtests. Estimate the possible range (magnitude) of each listedelemental error affecting the measured pressure refering to the solution provided online,
Sensitivity error(eK)
= (±0.0025)(500 cm H2O)= ± 0.75 cm H2O
= ± 0.00375 V
how do you get 0.75?? shoud it not be 1.25?
according the answer, for sensitivity
The transducer specified in Table 1.1 is chosen to measurea nominal pressure of 500 cm H2O. The ambient temperature is expected to vary between 18 ∘C and 25 ∘C duringtests. Estimate the possible range (magnitude) of each listedelemental error affecting the measured pressure.
How do I calculate the sensitivity error?
The solution that was given:
Sensitivity error(eK)
= (±0.0025)(500 cm H2O)= ± 0.75 cm H2O
= ± 0.00375 V
My Question is, how do you obtain 0.75 since 0.0025 x 500 gives 1.25!
please help. thanks
2. Consider the voltmeter calibration data in Table 1. The sensitivity of the pressure
sensor is 0.5 V/kPa (given by manufacturer) and the linear curve fit for the upscale
calibration is given as y = 0.9829 x + 0.1095 . Plot the data using a suitable scale.
a) Specify the percent maximum hysteresis based on full-scale range.
b) Referring to increasing input calibration, determined the sensitivity and
linearity errors.
Increasing input (mV)
Decreasing input (mV)
Y
Y
0.0
0.1
5.0
5.0
1.0
1.1
4.0
4.2
2.0
2.1
3.0
3.2
3.0
3.0
2.0
2.2
4.0
4.1
1.0
1.2
5.0
5.0
0.0
0.2
Table 1: Calibration results
Chapter 5 Solutions
Theory and Design for Mechanical Measurements
Ch. 5 - Prob. 5.1PCh. 5 - Explain what is meant by the terms “true value,"...Ch. 5 - SJ An official Olympics timekeeper once stated,...Ch. 5 - A tachometer has an analog display dial graduated...Ch. 5 - An automobile speedometer is graduated in 5-mph...Ch. 5 - An engineer reads the height of a mercury...Ch. 5 - Prob. 5.8PCh. 5 - Two resistors arc to be combined to form an...Ch. 5 - An equipment catalog boasts that a pressure...Ch. 5 - The shear modulus, G, of an alloy can be...
Ch. 5 - Heat transfer from a rod of diameter D immersed in...Ch. 5 - Estimate the design-stage uncertainty in...Ch. 5 - Explain the critical dift'erence(s) between a...Ch. 5 - A displacement transducer has the following...Ch. 5 - Prob. 5.18PCh. 5 - For a thin-walled pressure vessel of diameter D...Ch. 5 - Suppose a measured normal stress contains three...Ch. 5 - Prob. 5.31PCh. 5 - The area of a flat, rectangular parcel of land is...Ch. 5 - Estimate the random standard uncertainty in the...Ch. 5 - Estimate the uncertainty at 95% confidence in the...Ch. 5 - One use for a GPS rangefinder is on the golf...Ch. 5 - A pressure measuring system outputs a voltage that...Ch. 5 - The density of a metal composite is to be...Ch. 5 - A temperature measurement system is calibrated...Ch. 5 - The power usage of a strip heater is to be...Ch. 5 - The power usage of a DC strip heater can be...Ch. 5 - A thermocouple is a temperature sensor made of two...Ch. 5 - Time variations in a signal require that the...Ch. 5 - A hand-held infrared thermometer from a well-...Ch. 5 - The pressure in a large vessel is to be maintained...Ch. 5 - The cooling of a thermometer (c.g., Exs. 3.3 and...Ch. 5 - A J-type thermocouple monitors the temperature of...Ch. 5 - The density of air must be known to within 0.5%....Ch. 5 - In pneumatic conveying, solid particles such as...Ch. 5 - A step test is run to determine the time constant...Ch. 5 - The acceleration of a cart down a plane inclined...Ch. 5 - The heat flux in a reaction is estimated by Q =...Ch. 5 - A comparative test uses the relationship R =...Ch. 5 - A sensitive material is to be contained within a...Ch. 5 - A geometric stress concentration factor, K,, is...Ch. 5 - In Problem 5.9, we assumed that the errors in die...
Knowledge Booster
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, mechanical-engineering and related others by exploring similar questions and additional content below.Similar questions
- What are the five different classifications of Measurements? A force measurement system (weight scale) has the following specifications: Range 0 to 5,000 N Linearity error ±0.13% FSO Hysteresis ±0.14% FSO Sensitivity error ±0.20% of reading Thermal sensitivity error ±0.05% /oC of reading Zero drift ±0.03% /oC FSO Find the total uncertainty of reading if the output of weight scale reads 2,340 N at a temperature of 27 oC.arrow_forwardA pressure measurement system is stated as having the followingcharacteristics. Explain the significance of the terms:Range: 0 to 125 kPa and 0 to 2500 kPaAccuracy: ±1% of the displayed readingTemperature sensitivity: ±1% of the reading per o Carrow_forwardWe wish to measure a characteristic of a machined part whose nominal size is 50 mm at 20 °C. For this, two different instruments are used; an analog micrometer and a digital vernier caliper, both recently calibrated and with the same resolution (0.01 mm). The calibration certificate data of the analog micrometer and the digital vernier are shown in tables 5 and 6 respectively. In both cases the calibration uncertainty does not include the contribution of resolution uncertainty. Additionally, there are drift studies for both instruments and it is established that the drift of the micrometer is ± 0.002 mm while the drift of the vernier is ± 0.02 mm. The same number of measurements were performed with both instruments to avoid repeatability biases under controlled environmental conditions. Table 7 shows the results of the measurements of both instruments. The temperature of the piece was monitored with a digital thermometer whose calibration uncertainty is 0.1 °C, which already includes…arrow_forward
- 2- A pressure transducer is calibrated in an environment at a temperature of 20 °C and the following output readings (mV) are obtained for various input values P(kPa): P (kPa) ? 10 ? 20 ? 30 mV 12.1 24.2 | 36.3 | 48.4 | 60.5 72.6 a)Determine the sensitivity. b)lf the instrument is used in an environment at a temperature of 40 °C, it is noticed that the sensitivity became equals to the same value at 20 °C, but approximated to the nearest integer value, calculate the sensitivity drift per temperature change.arrow_forwardA displacement transducer has the following specifications: Linearity error:±0.25% reading Drift:±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V Output: dc 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltme- ter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Estimate an uncertainty in a nominal displacement of 2 cm at the design stage. Assume 95% confidence.arrow_forwardThe following resistance of a platinum resistance thermometer were measured at a range of temperature. Determine the measurement sensitivity of the instrument in milli Ohms/ oC. Resistance (mῼ) Tempreture (oC) 210k 110 217k 89 221k 68 228k 47arrow_forward
- A displacement transducer has the following specifications: Linearity error: ±0.25% reading Drift: ±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V dc Output: 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltmeter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Assume 95% confidence. The displacement transducer above is used in measuring the displacement of a body impacted by a mass. Twenty measurements are made, which yield X= 17.20mm Sx= 1.70mm Determine a best estimate for the mass displacement at 95% probability based on all available information.arrow_forwardA displacement transducer has the following specifications: Linearity error: ±0.25% reading Drift: ±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V dc Output: 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltmeter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Assume 95% confidence. The displacement transducer above is used in measuring the displacement of a body impacted by a mass. Twenty measurements are made, which yield Determine a best estimate for the mass displacement at 95% probability based on all available information.arrow_forwardA temperature measurement system is composed of a sensor and a readout device. The readout device has a claimed accuracy of 0.6 °C with a resolution of 0.1 °C. The sensor has an off-the-shelf limit of error of 0.5 °C. Estimate a design-stage uncertainty in the temperature indicated by this combination. A displacement transducer has the following specifications: Linearity error:±0.25% reading Drift:±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V Output: dc 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltme- ter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Estimate an uncertainty in a nominal displacement of 2 cm at the design stage. Assume 95% confidence.arrow_forward
- A Pitot tube measures a pressure difference Pa = Pt- Ps (dynamic = total - static) between the front and side of the Pitot tube. The dynamic pressure is given by the kinetic energy due to the airspeed v, allowing airspeed to be calculated from the measured pressure difference. (a) P₁ = ² pv² ⇒ v= When designing a Pitot tube, a manufacturer wants to make it precise enough to measure airspeed to within 0.5 % uncertainty, i.e. = 0.005, at sea level. Av p=(1.225 +0.0005) kg/m³ at sea level. (b) 2Pd Av Using error propagation, derive an expression for Simplify as much as possible! ΔΡΑ Pd Calculate the maximum permissible %-uncertainty in Pa, i.e. allow the Pitot tube to meet the manufacturer's requirements. Hint: Answer 1% x 100, that willarrow_forward(d) Based on the situation below, give one type of error faced by Ali's and give the suggestion how Ali's can make the reading become accurate and precise. Ali uses a vernier caliper to measure the diameter of 3 tubes. Then, the readings are compared to the actual values of the tubes as shown in Table 1: Tubes number Actual value Reading measurement 0.05 mm No tube Tube 1 5.05 mm 5 mm Tube 2 10.05 mm 10 mm Tube 3 12.05 mm 12 mm Table 1arrow_forward3. A consumer product company is formulating a new shampoo and is interested in foam height (in millimeters). Foam height is approximately normally distributed and has a standard deviation, 6, of 10 millimeters. The company wishes to test the following hypotheses using the results of n=25 samples: Ho: μ = 160, Ηi: μ < 160 The critical region to reject the null hypothesis is X < 155 %3D (a) Find the type I error probability a. (b) Find the probability of type II error if the true mean form height is 159 millimeters.arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Elements Of ElectromagneticsMechanical EngineeringISBN:9780190698614Author:Sadiku, Matthew N. O.Publisher:Oxford University PressMechanics of Materials (10th Edition)Mechanical EngineeringISBN:9780134319650Author:Russell C. HibbelerPublisher:PEARSONThermodynamics: An Engineering ApproachMechanical EngineeringISBN:9781259822674Author:Yunus A. Cengel Dr., Michael A. BolesPublisher:McGraw-Hill Education
- Control Systems EngineeringMechanical EngineeringISBN:9781118170519Author:Norman S. NisePublisher:WILEYMechanics of Materials (MindTap Course List)Mechanical EngineeringISBN:9781337093347Author:Barry J. Goodno, James M. GerePublisher:Cengage LearningEngineering Mechanics: StaticsMechanical EngineeringISBN:9781118807330Author:James L. Meriam, L. G. Kraige, J. N. BoltonPublisher:WILEY
Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press
Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON
Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education
Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY
Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning
Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY
How to balance a see saw using moments example problem; Author: Engineer4Free;https://www.youtube.com/watch?v=d7tX37j-iHU;License: Standard Youtube License