Interpretation:
The reason for the variation in the slit width of prism monochromator to obtain the constant effective bandwidth needs to be explained.
Concept introduction:
Monochromators are optical devices which transmit a wide range of available wavelengths, a mechanically selectable narrow wavelength band. These could be
Answer to Problem 7.1QAP
Over a certain spectral range, the output wavelength is continuously variable in case of prism monochromator and so, the slit width also must be varied.
Explanation of Solution
The wavelength, that is the output keeps changing continuously for a considerable spectral range in case of a prism monochromator while in case of a grating monochromator, the output wavelength remains same, as the radiation is dispersed in the component wavelength.
As the output wavelength keeps varying, the slit width is also required to be varied in case of a prism monochromator in order to keep the effective wavelength constant which is not needed in case of a grating monochromator.
Slit width must be varied to get constant effective wavelength in case of prism monochromator while it is almost same in case of grating monochromator.
Want to see more full solutions like this?
Chapter 7 Solutions
Principles of Instrumental Analysis
- The logarithm of the molar absorptivity of acetone in ethanol is 2.75 to 366 nm. Calculate the range of acetone concentrations that can be worked on if The transmittance percentage must be greater than 10% and less than 90% %, measured in a cell with an optical pitch of 1.50 cm.arrow_forwardExpress a wavelength of 442 nm as (a) a frequency. (b) a wavenumber.arrow_forwardExplain why graphite furnace atomisers provide enhanced sensitivity over flame atomisers in AAS measurements.arrow_forward
- Describe the similarities between fluorescence detectors in HPLC and chemiluminescence detectors in HPLC. What issues experienced with fluorescence detectors are resolved using chemiluminescence detectors?arrow_forwardWhy are wavlength maxima different for UV-Vis and Fluorescence?arrow_forwardWhen analyzing a selution component by UV-visible spectrophotometry. we prepare solutions with a range of concentrations by serial dilution. Why do we need to analyze a range of solutions? To allow multiple determinations of absorbance at different wavelengths and different concentrations To provide more than one determination of absorbance for a given concentration To see how absorbance varies with concentration To ensure that the linear relationship between absorbarce and concentrasion is velid for a range of concentrations.arrow_forward
- Distinguish between voltammetry and amperometryarrow_forwardWhat major impact does reducing film thickness of stationary phase in a wall-coated capillary column have on the separation efficiency (i.e., band broadening) in GC: Shorter tr (retention time) Lower C-term Lower resolution Higher A-term Higher B-termarrow_forwardExplain the difference of an operational principle between flame atomic absorption spectroscopy (FAAS) and flame atomic emission spectroscopy (AES). Predict the type of interference caused by the formation of metal ion-anion compounds which are difficult to be dissociated in flame.arrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningChemistry for Engineering StudentsChemistryISBN:9781285199023Author:Lawrence S. Brown, Tom HolmePublisher:Cengage Learning