Theory and Design for Mechanical Measurements
6th Edition
ISBN: 9781118881279
Author: Richard S. Figliola, Donald E. Beasley
Publisher: WILEY
expand_more
expand_more
format_list_bulleted
Concept explainers
Textbook Question
Chapter 1, Problem 1.54P
For the calibration data of Table 1.5, determine the static sensitivity of the system at (a) X = 5, (b) X = 10, and (c) X = 20. For which input values is the system more sensitive? Explain what this might mean in terms of a measurement and in terms of measurement errors.
Table 1.5 Calibration Data
X [cm] | r[V] | X [cm] | r[v) |
0.5 | 0.4 | 10.0 | 15.8 |
1.0 | 1.0 | 20.0 | 36.4 |
2.0 | 2.3 | 50.0 | 110.1 |
5.0 | 6.9 | 100.0 | 253.2 |
Expert Solution & Answer
Want to see the full answer?
Check out a sample textbook solutionStudents have asked these similar questions
Consider the voltmeter calibration data in the following table. Plot the data using a suitable scale.
Specify the percent maximum hysteresis based on full-scale range.
Increasing Input [mV]
Decreasing Input [mV]
Y
Y
0.0
0.1
5.0
5.0
1.0
1.1
4.0
4.2
2.0
2.1
3.0
3.2
3.0
3.0
2.0
2.2
4.0
4.1
1.0
1.2
5.0
5.0
0.0
0.2
2. Consider the voltmeter calibration data in Table 1. Plot the data using a suitable scale.
a) Specify the percent maximum hysteresis based on full-scale range.
b) Referring to increasing input calibration, determined the sensitivity and linearity errors.
Increasing input (mV)
Decreasing input (mV)
X
Y
Y
0.0
1.0
0.1
5.0
5.0
1.1
4.0
4.2
2.0
2.1
3.0
3.2
3.0
3.0
2.0
2.2
4.0
4.1
1.0
1.2
5.0
5.0
Table 1: Calibration results
0.0
0.2
b) Figure 1 shows the calibration data of a sensor. Determine the static sensitivity at the
input of X= 0, X= 5 and X= 10. For which input values is the system more sensitive?
Calibration Data
300
250
y = 0.875x1.055
200
150
100
50
20
40
60
80
100
120
Input value, cm
Figure 1
c) Suppose you found a dial thermometer in a stockroom. Discuss several methods by
which you might estimate random and systematic error in the thermometer? How
would you estimate its uncertainty?
Output value, V
Chapter 1 Solutions
Theory and Design for Mechanical Measurements
Ch. 1 - Prob. 1.1PCh. 1 - Prob. 1.2PCh. 1 - Prob. 1.3PCh. 1 - Prob. 1.4PCh. 1 - Prob. 1.5PCh. 1 - Discuss how the resolution of the display scale of...Ch. 1 - Prob. 1.7PCh. 1 - Prob. 1.8PCh. 1 - Prob. 1.9PCh. 1 - Prob. 1.10P
Ch. 1 - State the purpose of using randomization methods...Ch. 1 - Provide an example of repetition and replication...Ch. 1 - Develop a test plan that might be used to estimate...Ch. 1 - Develop a test plan that might be used to evaluate...Ch. 1 - A race engine shop has just completed two engines...Ch. 1 - A thermodynamics model assumes that a particular...Ch. 1 - Regarding the Mars Climate Orbiter spacecraft...Ch. 1 - A large batch of carefully made machine shafts can...Ch. 1 - Suggest an approach or approaches to estimate the...Ch. 1 - Suggest a test matrix to evaluate the wear...Ch. 1 - Figure 1.15 Orifice flow meter setup used for...Ch. 1 - The sale of motor fuel is an essential business in...Ch. 1 - Using either the ASME 19.5 or ISO 5167 test...Ch. 1 - A simple thermocouple circuit is formed using two...Ch. 1 - 1.25 A linear variable displacement transducer...Ch. 1 - For the LVDT calibration of the previous problem,...Ch. 1 - A manufacturer wants to quantify the expected...Ch. 1 - Prob. 1.28PCh. 1 - As described in a preceding problem, the...Ch. 1 - Light gates may be used to measure the speed of...Ch. 1 - You estimate your car’s fuel use by recording...Ch. 1 - When discussing concomitant methods, we used the...Ch. 1 - Prob. 1.33PCh. 1 - For the strain gauge calibration of the previous...Ch. 1 - The acceleration of a cart down a plane inclined...Ch. 1 - In general, what is meant by the term “standard”?...Ch. 1 - A common scenario: An engineer has two pencil-...Ch. 1 - Explain the potential differences in the following...Ch. 1 - Research the following test standards and codes....Ch. 1 - A hotel chain based in the United States contracts...Ch. 1 - Test code ASTM 558-13 allows for the comparison of...Ch. 1 - Suggest a reasonable number of significant digits...Ch. 1 - Using spreadsheet software (such as Microsoft...Ch. 1 - Prob. 1.44PCh. 1 - Round the following numbers to 3 significant...Ch. 1 - Express the result, rounding to an appropriate num...Ch. 1 - Express the result by rounding to an appropriate...Ch. 1 - A car’s speed is determined by the time it takes...Ch. 1 - How much error could you tolerate in (1) book...Ch. 1 - Apply the guidelines to determine the number of...Ch. 1 - Using a tape measure having 1 mm graduations, the...Ch. 1 - Show how the following functions can be...Ch. 1 - Prob. 1.53PCh. 1 - For the calibration data of Table 1.5, determine...Ch. 1 - Prob. 1.55PCh. 1 - Each of the following equations can be represented...Ch. 1 - Plot y = 10e“° 5x volts on in semilog format (use...Ch. 1 - Prob. 1.58PCh. 1 - Prob. 1.59P
Knowledge Booster
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, mechanical-engineering and related others by exploring similar questions and additional content below.Similar questions
- 4:07 3G l. Moodle A bbb2.du.edu.om EE ...Session (15-3- Said Grami O College of Engineering Vision for the Future Question 3 During a step function calibration, a first-order instrument is exposed to a step change of 200 units. 1- If after 1.5 s the instrument indicates 140 units, find the instrument time constant. 2- find the error in the indicated value after 2 s S. 3- Find the raise time at 90 % of the system. y(0)=0 units; K=1 unit/unit. Question 4 A first order instrument with a time constant of 2 seconds is to be used to measure a periodic input. If a dynamic error of ±2% can be tolerated, 1- determine the maximum frequency of periodic input that can be measured. 2- What is the associated time lag (in seconds) at that frequency? MECH374: Instrumentation and Measurement Said Grami Practicearrow_forwardLab 2-Measurement Asynch - Tagged.pdf Page 4 of 7 ? Part I: Taking Measurements & Estimating Uncertainties for a single measurement www.stefanelli.eng.br The mass of the object is_ 0 i Parts on a tripie peam palance 0 0 10 20 30 1 100 2 3 40 200 4 +/- 50 60 70 5 300 7 400 80 Qv Search 8 90 9 500 100 9 10 g www.stefanelli.eng.brarrow_forwardA temperature measurement system has the following specifications: -128 to 781°C Range Linearity error 0.29% FSO Hysteresis error 0.12% FSO Sensitivity error 0.04% FSO Zero drift 0.32% FSO FSO stands for "Full Scale Output". Estimate the overall instrument uncertainty for this system based on the available information. Use the maximum possible output range over the FSO in your computations.arrow_forward
- A Pitot tube measures a pressure difference Pa = Pt- Ps (dynamic = total - static) between the front and side of the Pitot tube. The dynamic pressure is given by the kinetic energy due to the airspeed v, allowing airspeed to be calculated from the measured pressure difference. (a) P₁ = ² pv² ⇒ v= When designing a Pitot tube, a manufacturer wants to make it precise enough to measure airspeed to within 0.5 % uncertainty, i.e. = 0.005, at sea level. Av p=(1.225 +0.0005) kg/m³ at sea level. (b) 2Pd Av Using error propagation, derive an expression for Simplify as much as possible! ΔΡΑ Pd Calculate the maximum permissible %-uncertainty in Pa, i.e. allow the Pitot tube to meet the manufacturer's requirements. Hint: Answer 1% x 100, that willarrow_forwardRecord the dimensions of the known (calibration) block using the caliper and dial gauge on the table below. Indicate the units of each measurement. Calculate the average length of each side of the block. Dimension Caliper (Units) 0.995 1.455 0.985 Ruler(in) A: 0.9 B: 1.5 C: 0.9 A B C Dimension A B Instrument Use the average dimensions (see Problem 2a) of the known block to calibrate the LVDT at your workstation. Record the voltage on the table below: LVDT Offset: 0.556 (Do not include the offset value in your average dimensions) C Ave Dimension (Units) (Dial Gauge) 0.997 1.659 0.949 0.964 in 1.538 in 0.945 in oltage Average Dimension 1.244 volt 1.994 1.28 0.964 in 1.538 in 0.945 inarrow_forwardA constant volume gas thermometer is used to determine the temperature of an unknown fluid. Pressure data for thermometer in the unknown bath (P) and a Triple Point Cell (P) are given below. P[Torr] 100.0 P [Torr] 127.9 200.0 256.5 300.0 385.8 400.0 516.0 What is the temperature of the unknown fluid bath? explain and show all work please and write clearlyarrow_forward
- A temperature measurement system is composed of a sensor and a readout device. The readout device has a claimed accuracy of 0.6 °C with a resolution of 0.1 °C. The sensor has an off-the-shelf limit of error of 0.5 °C. Estimate a design-stage uncertainty in the temperature indicated by this combination. A displacement transducer has the following specifications: Linearity error:±0.25% reading Drift:±0.05%/°C reading Sensitivity error: ±0.25% reading Excitation: 10-25 V Output: dc 0-5Vdc Range: 0-5 cm The transducer output is to be indicated on a voltme- ter having a stated accuracy of 0:1% reading with a resolution of 10 μV. The system is to be used at room temperature, which can vary by 10 °C. Estimate an uncertainty in a nominal displacement of 2 cm at the design stage. Assume 95% confidence.arrow_forwardc) In a particular process, the output result of force is a function of the displacement by F(kN) = = x-¹ + 2x where x is in [cm] Determine the sensitivity of force to displacement at 1 cm by dithering. Compare this to the analytical result based on the partial derivative.arrow_forward2- A pressure transducer is calibrated in an environment at a temperature of 20 °C and the following output readings (mV) are obtained for various input values P(kPa): P (kPa) ? 10 ? 20 ? 30 mV 12.1 24.2 | 36.3 | 48.4 | 60.5 72.6 a)Determine the sensitivity. b)lf the instrument is used in an environment at a temperature of 40 °C, it is noticed that the sensitivity became equals to the same value at 20 °C, but approximated to the nearest integer value, calculate the sensitivity drift per temperature change.arrow_forward
- . 2) A following figure and table display creep test data obtained from a uniaxial specimen having an initial gauge length of 2 cm. Calculate the steady-state creep rate. Length |(cm) 2.3 Time | (h) 2.25 2.004 2.2 100 2.010 200 2.020 400 2.030 1000 2000 2.045 2.075 2.1 2.05 4000 2.135 6000 2.193 7000 2.230 1.95 8000 2.300 1000 2000 3000 4000 5000 6000 7000 8000 Time (h) (rupture) | Length between the gauge marks (cm) 2.arrow_forwardPartII Say you've deployed a robot that can can measure electric field due to specified charge at the specified location with good accuracy. And you got these results from the robot: Electric field due to q2 at point p1 is (-1667.0) i + (18340.0) j Electric field due to q3 at point p1 is (-2496.0) i + (–1628.0) j Electric field due to q1 at point p2 is (-6311.0) î + (2470.0) } Electric field due to q2 at point p2 is (-11585.0) î + (–6951.0) ĵ Electric field due to q3 at point p2 is (4657.0) î + (11310.0) } c) Find the net Electric field at pj .arrow_forwardA thermocouple was used to measure the temperature. Over the range of temperatures expected in the experiment, a linear regression analysis was performed for the temperature as a function of voltage. The best-fit equation was T = (-0.008563 + 220*V), where V is the thermocouple voltage. The voltage was fed into a 11-bit A/D converter with a voltage range of -8 V to 8 V. Determine the measurement resolution (in °C) due to the ADC? O 0.851 None O 3.113 1.710 0.393 4.960 1.144 O 3.797arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Elements Of ElectromagneticsMechanical EngineeringISBN:9780190698614Author:Sadiku, Matthew N. O.Publisher:Oxford University PressMechanics of Materials (10th Edition)Mechanical EngineeringISBN:9780134319650Author:Russell C. HibbelerPublisher:PEARSONThermodynamics: An Engineering ApproachMechanical EngineeringISBN:9781259822674Author:Yunus A. Cengel Dr., Michael A. BolesPublisher:McGraw-Hill Education
- Control Systems EngineeringMechanical EngineeringISBN:9781118170519Author:Norman S. NisePublisher:WILEYMechanics of Materials (MindTap Course List)Mechanical EngineeringISBN:9781337093347Author:Barry J. Goodno, James M. GerePublisher:Cengage LearningEngineering Mechanics: StaticsMechanical EngineeringISBN:9781118807330Author:James L. Meriam, L. G. Kraige, J. N. BoltonPublisher:WILEY
Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press
Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON
Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education
Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY
Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning
Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY
What is Metrology in Mechanical Engineering? | Terminologies & Measurement; Author: GaugeHow;https://www.youtube.com/watch?v=_KhMhFRehy8;License: Standard YouTube License, CC-BY