In the lab, you have setup a thermocouple and have used a thermistor along with an ice
bath and water at various temperatures (confirmed with the thermistor) up to 100
degrees Celsius for calibration. The calibration data is shown in the table below and
the full-scale output range is 0-5 mV.
You note that there is scatter in your data; however, you must use a linear curve fit to
efficiently process the measurements during an automated temperature measurement
process.
Question 1
100%
Question 2
100%
Question 3
100%
Question 4
100%
Question 5
100%
Question 6
100%
mV
The slope of your linear calibration curve for the thermocouple is 0.0334 °C with an
offset of -0.07 mV.
Question 7
100%
Question 8
100%
What is the maximum expected linearity error as a percentage of the full-scale output?
Question 9
0%
Summary
-0.08
Thermocouple Calibration Data
Temperature (°C) Voltage (in mV)
0
20
20
40
40
60
60
60
80
96
90
0.587
1.314
1.901
2.528
2.782
100
3.055
LIT
MacBook Pro
<6
%
905
34
E
R
T
&
27
Y
* 8
כ
6
O
expand button
Transcribed Image Text:Maul Dulde Questio119 819 PREV NEXT In the lab, you have setup a thermocouple and have used a thermistor along with an ice bath and water at various temperatures (confirmed with the thermistor) up to 100 degrees Celsius for calibration. The calibration data is shown in the table below and the full-scale output range is 0-5 mV. You note that there is scatter in your data; however, you must use a linear curve fit to efficiently process the measurements during an automated temperature measurement process. Question 1 100% Question 2 100% Question 3 100% Question 4 100% Question 5 100% Question 6 100% mV The slope of your linear calibration curve for the thermocouple is 0.0334 °C with an offset of -0.07 mV. Question 7 100% Question 8 100% What is the maximum expected linearity error as a percentage of the full-scale output? Question 9 0% Summary -0.08 Thermocouple Calibration Data Temperature (°C) Voltage (in mV) 0 20 20 40 40 60 60 60 80 96 90 0.587 1.314 1.901 2.528 2.782 100 3.055 LIT MacBook Pro <6 % 905 34 E R T & 27 Y * 8 כ 6 O
Thermocouple Calibration Data
Temperature (°C)
Voltage (in mV)
0
-0.08
20
0.587
40
1.314
60
1.901
80
2.528
90
2.782
100
3.055
The
Seera: no
MacB
expand button
Transcribed Image Text:Thermocouple Calibration Data Temperature (°C) Voltage (in mV) 0 -0.08 20 0.587 40 1.314 60 1.901 80 2.528 90 2.782 100 3.055 The Seera: no MacB
Expert Solution
Check Mark
Knowledge Booster
Background pattern image
Similar questions
SEE MORE QUESTIONS
Recommended textbooks for you
Text book image
Introductory Circuit Analysis (13th Edition)
Electrical Engineering
ISBN:9780133923605
Author:Robert L. Boylestad
Publisher:PEARSON
Text book image
Delmar's Standard Textbook Of Electricity
Electrical Engineering
ISBN:9781337900348
Author:Stephen L. Herman
Publisher:Cengage Learning
Text book image
Programmable Logic Controllers
Electrical Engineering
ISBN:9780073373843
Author:Frank D. Petruzella
Publisher:McGraw-Hill Education
Text book image
Fundamentals of Electric Circuits
Electrical Engineering
ISBN:9780078028229
Author:Charles K Alexander, Matthew Sadiku
Publisher:McGraw-Hill Education
Text book image
Electric Circuits. (11th Edition)
Electrical Engineering
ISBN:9780134746968
Author:James W. Nilsson, Susan Riedel
Publisher:PEARSON
Text book image
Engineering Electromagnetics
Electrical Engineering
ISBN:9780078028151
Author:Hayt, William H. (william Hart), Jr, BUCK, John A.
Publisher:Mcgraw-hill Education,