An analyst was measuring the performance of a HPLC column. The test solute was Toluene (5 µL). The column was a 4.6 mm i.d. column, 150 mm in length and packed with 5 µm C18 particles. The mobile phase was 40/60 water/methanol and the flow rate was 0.80 mL/min. The void time (to) was 1.86 minutes The retention time of the toluene sample was 14.50 min. The standard deviation of the peak was 7.0542 (s). 2a. The mobile phase velocity was?

Principles of Instrumental Analysis
7th Edition
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Chapter26: An Introduction To Chromatographic Separations
Section: Chapter Questions
Problem 26.14QAP
icon
Related questions
Question
An analyst was measuring the performance of a HPLC column. The test solute was Toluene (5 µL). The
column was a 4.6 mm i.d. column, 150 mm in length and packed with 5 µm C18 particles. The mobile
phase was 40/60 water/methanol and the flow rate was 0.80 mL/min. The void time (to) was 1.86
minutes
The retention time of the toluene sample was 14.50 min.
The standard deviation of the peak was 7.0542 (s).
2a. The mobile phase velocity was?
0.0806 mm/s
0.1344 mm/s
1.344 mm/s
1.344 x 10³ mm/s
(i)
(ii)
(iv)
2b. The number of theoretical plates was?
(i)
(ii)
(iii)
(iv)
2c. The height equivalent to a theoretical plate was?
1.041x10³ m
1.015x105 m
9.862x10-6 m
1.186x10¹² m
(i)
(ii)
(iii)
(iv)
4802
15210
24705
9848
2d. The reduced plate height was?
(i)
2.03
(ii)
1.97
(iii)
7.55
(iv)
2.08
2e. The column was?
(i)
(ii)
Extremely well packed
Well packed
Poorly packed
(iii)
(iv) Insufficient information to decide about the quality of the column
Transcribed Image Text:An analyst was measuring the performance of a HPLC column. The test solute was Toluene (5 µL). The column was a 4.6 mm i.d. column, 150 mm in length and packed with 5 µm C18 particles. The mobile phase was 40/60 water/methanol and the flow rate was 0.80 mL/min. The void time (to) was 1.86 minutes The retention time of the toluene sample was 14.50 min. The standard deviation of the peak was 7.0542 (s). 2a. The mobile phase velocity was? 0.0806 mm/s 0.1344 mm/s 1.344 mm/s 1.344 x 10³ mm/s (i) (ii) (iv) 2b. The number of theoretical plates was? (i) (ii) (iii) (iv) 2c. The height equivalent to a theoretical plate was? 1.041x10³ m 1.015x105 m 9.862x10-6 m 1.186x10¹² m (i) (ii) (iii) (iv) 4802 15210 24705 9848 2d. The reduced plate height was? (i) 2.03 (ii) 1.97 (iii) 7.55 (iv) 2.08 2e. The column was? (i) (ii) Extremely well packed Well packed Poorly packed (iii) (iv) Insufficient information to decide about the quality of the column
Expert Solution
steps

Step by step

Solved in 4 steps

Blurred answer
Knowledge Booster
Analytical Separations
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, chemistry and related others by exploring similar questions and additional content below.
Similar questions
  • SEE MORE QUESTIONS
Recommended textbooks for you
Principles of Instrumental Analysis
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
Fundamentals Of Analytical Chemistry
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:
9781285640686
Author:
Skoog
Publisher:
Cengage
Appl Of Ms Excel In Analytical Chemistry
Appl Of Ms Excel In Analytical Chemistry
Chemistry
ISBN:
9781285686691
Author:
Crouch
Publisher:
Cengage